dc.description.abstract
Hintergrund: Der Erfolg der perkutanen transluminalen Angioplastie in
peripheren Arterien wird durch hohe Restenoseraten limitiert. Ziel der
vorliegenden Studie war es, die Sicherheit und Wirksamkeit Paclitaxel-
beschichteter Angioplastie-Ballonkatheter zur Behandlung arteriosklerotisch
bedingter Stenosen und Okklusionen der femoropoplitealen Arterien zu
untersuchen. Methoden: Insgesamt wurden 87 Patienten in diese prospektive
Multicenterstudie eingeschlossen. Es erfolgte randomisiert eine Behandlung mit
unbeschichteten Ballonkathetern (N = 42) oder Paclitaxel-beschichteten
Ballonkathetern (N = 45). Primärer Endpunkt war der Late Lumen Loss (LLL) nach
sechs Monaten. Ergebnisse: Bei Studieneinschluss wiesen beide Gruppen ähnliche
demographische Daten und Läsionscharakteristika auf, z. B. durchschnittliche
Läsionslänge (5,7 cm versus 6,1 cm), Stenosegrad (84% ± 11% versus 84% ± 16%)
und Anteil an Restenosen (36% versus 33%). Zum 6-Monats-Follow-up wurde eine
Angiographie bei 31 der 45 Patienten in der Paclitaxel-Gruppe und 34 der 42
Patienten in der Kontrollgruppe durchgeführt. Im Vergleich zur Kontrollgruppe
zeigten Patienten, welche mit Paclitaxel-beschichteten Ballonkathetern
behandelt wurden, einen geringeren LLL (1,0 ± 1,1 mm versus 0,5 ± 1,1 mm; p =
0,031) und eine niedrigere binäre Restenoserate (47,1% versus 19,4%; p =
0,035). Die Revaskularisationsrate in der Studienläsion zum 6-Monats-Follow-up
betrug 3 von 45 (6,7%) in der Paclitaxel-Gruppe und 14 von 42 (33,3%) in der
Kontrollgruppe (p = 0,0024). Patienten, die mit Paclitaxel-beschichteten
Ballonkathetern behandelt wurden, wiesen eine stärkere Verbesserung des
Rutherford-Stadiums im Vergleich zur Kontrollgruppe auf (p = 0,045). Bezüglich
des Knöchel-Arm-Index zeigte sich kein Unterschied zwischen den beiden
Gruppen. Unerwünschte Ereignisse, welche auf die Paclitaxel-Beschichtung
zurück zu führen waren, traten nicht auf. Fazit: Die Anwendung Paclitaxel-
beschichteter Angioplastie-Ballonkatheter bei arteriosklerotisch bedingten
Stenosen und Okklusionen der femoropoplitealen Arterien führte zu einer
signifikanten Senkung LLL und der binären Restenoserate. Es wurden keine
unerwünschten Ereignisse im Zusammenhang mit der Paclitaxel-Beschichtung
beobachtet.
de
dc.description.abstract
Background: The success of percutaneous transluminal angioplasty in peripheral
arteries is limited by high restenosis rates. The aim of the present study was
to evaluate the safety and efficacy of paclitaxel-coated angioplasty balloons
for the treatment of femoropopliteal peripheral artery disease. Methods: A
total of 87 patients were enrolled in this prospective multicenter study and
randomized to treatment with either conventional uncoated balloons (n = 42) or
paclitaxel-coated balloon catheters (n = 45). The primary end point was late
lumen loss at 6 month. Results: At baseline demographics and lesion
characteristics were similar in both groups, e.g. mean lesion length (5.7 cm
versus 6.1 cm), degree of stenosis (84% ± 11% versus 84% ± 16%) and proportion
of restenotic lesions (36% versus 33%). Angiography at 6-month follow-up was
performed in 31 of 45 patients in the paclitaxel-group and 34 of 42 patients
in the control group. Compared to the control group, the group treated with
paclitaxel-coated balloons exhibited less late lumen loss (1.0 ± 1.1 mm versus
0.5 ± 1.1 mm; p = 0.031) and fewer binary restenoses rate (47.1% versus 19.4%;
p = 0.035). The rate of revascularisation of target lesions at 6 month was 3
of 45 (6.7%) in the paclitaxel-coated balloon group and 14 of 42 (33.3%) in
the control group (p=0.0024). Patients treated with the paclitaxel-coated
balloons showed a greater improvement in Rutherford class compared to control
subjects (p = 0.045). The improvement in ankle brachial index was not
different between the two groups. There were no paclitaxel-related adverse
events reported. Conclusion: Use of paclitaxel-coated balloon catheters for
the treatment of femoropopliteal peripheral artery disease was associated with
significant reductions in late lumen loss and binary restenosis at 6 month. No
adverse events related to the paclitaxel-coating were observed.
en
dc.rights.uri
http://www.fu-berlin.de/sites/refubium/rechtliches/Nutzungsbedingungen
dc.subject
drug-eluting balloons
dc.subject
peripheral arterial disease
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit
dc.title
Reokklusionsprophylaxe mittels Paclitaxel (Taxol®)-beschichteter Angioplastie-
Ballonkatheter bei arteriosklerotisch bedingten Stenosen und Okklusionen der
femoropoplitealen Arterien
dc.contributor.contact
Bianka.Reinkensmeier@gmx.de
dc.contributor.firstReferee
N.N.
dc.contributor.furtherReferee
N.N.
dc.date.accepted
2016-06-05
dc.identifier.urn
urn:nbn:de:kobv:188-fudissthesis000000101828-6
dc.title.translated
Reocclusion prophylaxis using paclitaxel (Taxol®)-coated balloon angioplasty
catheters at arteriosclerotic stenosis and occlusions of the femoropopliteal
arteries
en
refubium.affiliation
Charité - Universitätsmedizin Berlin
de
refubium.mycore.fudocsId
FUDISS_thesis_000000101828
refubium.mycore.derivateId
FUDISS_derivate_000000019042
dcterms.accessRights.dnb