dc.description.abstract
The utilization of epidural electrodes in the preoperative evaluation of intractable epilepsy is a valuable but underrepresented tool. In recent years, we have adapted the use of cylindrical epidural 1-contact electrodes (1-CE) instead of Peg electrodes. 1-CEs are more versatile since their explantation is a possible bedside procedure. Here we report our experience with 1-CEs as well as associated technical nuances. This retrospective analysis included 56 patients with intractable epilepsy who underwent epidural electrode placement for presurgical evaluation at the Department of Neurosurgery at the Charité University Hospital from September 2011 to July 2021. The median age at surgery was 36.3 years (range: 18–87), with 30 (53.6%) female and 26 (46.4%) male patients. Overall, 507 electrodes were implanted: 93 Fo electrodes, 33 depth electrodes, and 381 epidural electrodes, with a mean total surgical time of 100.5 ± 38 min and 11.8 ± 5 min per electrode. There was a total number of 24 complications in 21 patients (8 Fo electrode dislocations, 6 CSF leaks, 6 epidural electrode dislocations or malfunction, 3 wound infections, and 2 hemorrhages); 11 of these required revision surgery. The relative electrode complication rates were 3/222 (1.4%) in Peg electrodes and 3/159 (1.9%) in 1-CE. In summary, epidural recording via 1-CE is technically feasible, harbours an acceptable complication rate, and adequately replaces Peg electrodes.
en
dc.rights.uri
https://creativecommons.org/licenses/by/4.0/
dc.subject
epidural electrodes
en
dc.subject
Peg electrodes
en
dc.subject
invasive diagnostics
en
dc.subject
Fo electrodes
en
dc.subject
depth electrodes
en
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit::610 Medizin und Gesundheit
dc.title
Utilization of Epidural Electrodes as a Diagnostic Tool in Intractable Epilepsy—A Technical Note
dc.type
Wissenschaftlicher Artikel
dcterms.bibliographicCitation.articlenumber
397
dcterms.bibliographicCitation.doi
10.3390/mi13030397
dcterms.bibliographicCitation.journaltitle
Micromachines
dcterms.bibliographicCitation.number
3
dcterms.bibliographicCitation.originalpublishername
MDPI AG
dcterms.bibliographicCitation.volume
13
refubium.affiliation
Charité - Universitätsmedizin Berlin
refubium.resourceType.isindependentpub
no
dcterms.accessRights.openaire
open access
dcterms.bibliographicCitation.pmid
35334689
dcterms.isPartOf.eissn
2072-666X