Im Rahmen der vorliegenden Dissertation wurden handelsübliche Implantatsysteme
mit unterschiedlich gestalteten Implantat-Abutment-Verbindungen auf
Positionsabweichungen des Abutments bei wiederholter manueller De- und
Remontage der Implantatkomponenten untersucht. Material und Methode Zu diesem
Zwecke wurden je sechs Implantatkörper von fünf Implantatsystemen (Straumann,
SteriOss, Camlog, Astra Tech, Replace Select) in unterschiedlicher Angulation
in einem Metallblock fixiert. Drei Testpersonen lösten und befestigten die
Abutments, an welchen ein standardisierter Prüfkörper angebracht war,
anschließend in 20facher Wiederholung mithilfe des systembezüglichen
Schraubendrehers. Mittels Koordinatenmessmaschine (Messungenauigkeit: 0,7 μm
bzw. 0,004 °) wurden in Bezug auf ein Referenzkoordinatensystem
Positionsabweichungen im dreidimensionalen Raum ermittelt. Rotatorisch
bedingte Positions- sowie Höhenänderungen und Verkippungen der betreffenden
Implantate wurden untersucht. Die statistische Auswertung erfolgte mithilfe
einer parameterfreien Varianzanalyse nach Brunner. Ergebnisse Die getesteten
Verbindungen zeigten eine von 0,002 ° - 4,8 ° reichende Rotationsfreiheit.
Eines der untersuchten Implantatsysteme (Camlog) verhielt sich signifikant
unterschiedlich zu allen weiteren getesteten Verbindungen (p<0.001). Vertikale
Lageveränderungen von 1 bis 83 μm resultierten nach manueller Repositionierung
des Abutments; bezüglich dieser Positionsänderungen ergaben sich statistisch
signifikante Unterschiede zwischen Implantatsystemen mit horizontalem Anschlag
und konischen Implantat-Abutment-Verbindungen. Verkippende Momente konnten nur
in geringem Ausmaß beobachtet werden (0.00 ° - 0.28 °). Ein statistisch
signifikanter Einfluss des Implantatsystems oder der ausführenden Testperson
konnte nicht ermittelt werden. Schlussfolgerung Im Rahmen der vorliegenden
Untersuchung erwiesen sich einzelne Implantat-Abutment-Verbindungen
rotationsunsicherer und auch bezüglich vertikaler Lagestabilität weniger exakt
als angepriesen. Die ermittelten, nach manueller De- und Remontage
resultierenden, dreidimensionalen Lageabweichungen des Abutments nach
Repositionierung auf dem Implantat können in Abhängigkeit vom Implantattyp
Passungsungenauigkeiten der implantatgetragenen Suprakonstruktion bedingen.
de
dc.description.abstract
In this experiment the precision of manually repositioned abutments in five
implant systems with various implant-abutment interfaces was evaluated.
Material und Method Of these five implant systems (Straumann, Astra Tech,
Replace Select, Camlog, SteriOss) six angled (0°, 5°, 15°) implants each were
fixated in a prefabricated metal block. Three persons with differing
experience and knowledge of the hypothesis placed and removed a prefabricated
abutment with anti-rotational features 20 times on each of the six implants
per system using the system-specific screwdriver. A precision Video check
analysis system (precision 0,7 μm and 0,004°) detected the deviation of
position of a standardized test body attached to the abutment and it was
measured three-dimensional in a coordinate system. Rotational and vertical
changes as well as changes in angulation were monitored and statistically
evaluated using the non-parametric analysis of variance of Brunner. Results
The tested interfaces showed a deviation of 0.002 ° - 4.8 ° in rotation, one
implant system (Camlog) showed a significant difference in rotational accuracy
to all other tested systems (P<0.001). Vertical alterations ranged from 1 to
83 μm in height. Statistically significant differences exist between butt-
joint and conical implant-abutment connections. Straumann and Astra Tech
implants showed a significant difference to all other implant systems included
in the evaluation. Canting discrepancies (range 0.00 ° - 0.28 °) were minor
distinctive and not influenced by implant systems or test persons. Conclusion
The investigation showed that some commercially available implant-abutment
connections are less precise than promoted. This three-dimensional inaccuracy
of repositioned abutments, in dependence of the system used, may contribute to
the problems encountered with implant-retained restorations.
en
dc.rights.uri
http://www.fu-berlin.de/sites/refubium/rechtliches/Nutzungsbedingungen
dc.subject
implant-abutment-connection
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit::610 Medizin und Gesundheit
dc.title
Experimentelle Untersuchung der Positionsstabilität des Implantat-Abutment-
Komplexes unterschiedlicher Systeme bezüglich rotatorischer, vertikaler und
verkippender Lageänderungen bei wiederholter manueller De- und Remontage
dc.contributor.firstReferee
Prof. Dr. Dr. M. Klein
dc.contributor.furtherReferee
Prof. Dr. Dr. M. Herzog
dc.contributor.furtherReferee
Prof. Dr. Dr. B. Hell
dc.date.accepted
2008-02-22
dc.date.embargoEnd
2008-03-27
dc.identifier.urn
urn:nbn:de:kobv:188-fudissthesis000000003555-0
dc.title.translated
Effects of repeated manual disassembly and reassembly on the position
stability of various implant-abutment complexes
en
refubium.affiliation
Charité - Universitätsmedizin Berlin
de
refubium.mycore.fudocsId
FUDISS_thesis_000000003555
refubium.mycore.transfer
http://www.diss.fu-berlin.de/2008/47/
refubium.mycore.derivateId
FUDISS_derivate_000000003555
dcterms.accessRights.dnb
free
dcterms.accessRights.openaire
open access