dc.description.abstract
Fragestellung: Ziel der Untersuchung war, die Aussagekraft und Verlässlichkeit
der digitalen Volumentomographie bei der Darstellung filigraner parodontaler
Strukturen zu analysieren. Material und Methode: Fünf knöcherne menschliche
Schädel wurden bei 85 kV und 35/42 mAs volumentomographisch untersucht
(GALILEOS comfort, Sirona Dental GmbH, Bensheim, Deutschland). An allen Zähnen
(0,3 mm Voxelgröße) wurden die vestibulären Knochenverhältnisse bezüglich
Knochenhöhe, Dehiszenzen und Fenestrationen untersucht und vermessen.
Entsprechende Messungen mit einem Messschieber an den Schädeln dienten als
Referenz. Nach Präparation einzelner röntgenologisch nicht sichtbarer
Knochenlamellen erfolgte die stereomikroskopische Bestimmung der Knochendicke.
Außerdem wurde der Einfluss der Bildauflösung, Belichtungszeit, Zahnregion,
Defektart und Schmelz-Zement-Grenze sowie der Bildeinstellungen Kontrast und
Helligkeit auf die Messgenauigkeit analysiert und die Darstellung der
Wurzeloberfläche im Röntgenbild auf einen Zusammenhang mit den Ergebnissen hin
überprüft. Ergebnisse: Die Messungen zeigten eine durchschnittlich zu geringe
Darstellung der Knochenhöhe um 0,13 mm (35 mAs, p=0,000) bzw. 0,11 mm (42 mAs,
p=0,005). Die mittleren Abweichungen von der Referenz betrugen 0,26 bzw. 0,28
mm. Weder die Erkennbarkeit der Schmelz-Zement-Grenze noch die Veränderung von
Kontrast und Helligkeit hatten einen Einfluss auf das Messergebnis. Auch
zwischen den beiden Belichtungszeiten fand sich kein statistisch erkennbarer
Unterschied, es konnte jedoch eine Tendenz zur besseren Messgenauigkeit bei 42
mAs festgestellt werden. Eine tendenziell schlechtere Messgenauigkeit zeigte
sich in der Frontzahnregion (vor allem im Unterkiefer) sowie bei klinisch
vorhandenen Dehiszenzen, die in Einzelfällen deutlich überschätzt wurden. Die
Erkennung von Dehiszenzen im Röntgenbild war insgesamt zuverlässig
(Sensitivität 97 bzw. 94 %, Spezifität 98 %), im Gegensatz zu den
Fenestrationen, die zumeist vergrößert dargestellt oder, vor allem im
Oberkiefer, falsch-positiv diagnostiziert wurden (Sensitivität 100 %,
Spezifität 86 %). Die Erkennbarkeitsgrenze der Knochenlamellen lag bei einer
Knochendicke von 0,39 mm (0,3 mm Voxelgröße) und 0,22 mm (0,15 mm
Kantenlänge). Eine Voxelgröße von 0,15 mm lieferte im Vergleich zu 0,3 mm
signifikant bessere Ergebnisse. Bei abschließender Betrachtung der Ergebnisse
fiel auf, dass eine subjektiv unscharfe Darstellung der Wurzeloberfläche im
Bereich von röntgenologischen Dehiszenzen oder Fenestrationen ein Hinweis auf
eine falsch-positive Diagnose des Knochendefekts sein kann. Schlussfolgerung:
Die digitale Volumentomographie kann grundsätzlich als geeignetes Verfahren
für die präzise und überlagerungsfreie Darstellung parodontaler Strukturen
betrachtet werden. Eine weitestgehend ausreichend präzise Darstellung der
Knochenmorphologie bei einer Voxelgröße von 0,3 mm konnte festgestellt werden.
Probleme ergeben sich, wenn sich die Dicke der Knochenstrukturen im Bereich
der Auflösungsgrenze des Gerätes liegt. Hierbei kann es zur falsch-positiven
Diagnose von Knochendefekten kommen, daher sollten beim Einsatz der DVT für
spezifische parodontaldiagnostische Fragestellungen hochauflösende Geräte
gewählt werden.
de
dc.description.abstract
Introduction: Aim of this study was to investigate accuracy and reliability of
cone-beam computed tomography for the imaging of thin vestibular periodontal
structures. Materials and methods: Five human skulls were scanned by CBCT
(GALILEOS comfort, Sirona Dental GmbH, Bensheim, Deutschland) at 85 kV and
35/42 mAs. Linear measurements were performed on each tooth (0,3 mm voxel
size), evaluating periodontal bone height, bony dehiscences and fenestrations.
Results were compared to data measured by caliper on the skulls. After
preparation of several lamellas of undetectable bone their thickness was
examined by stereomicroscopy. In addition, influence of voxel size, exposure
time, type of defect, tooth area, detection of the cemento-enamel-junction and
variations of contrast / lightness on measuring accuracy were evaluated.
Results: Evaluations by CBCT showed a significant underestimation of buccal
bone height (0,13 mm for 35 mAs, p=0,000; 0,11 mm for 42 mAs, p=0,005). Mean
deviations to reference were 0,26 mm and 0,28 mm, respectively. There was
neither a significant influence of perceptibility of the cemento-enamel-
junction, nor of the variation of contrast and brightness. Moreover, the two
different exposure times showed no statistical difference, however, a tendency
for more accurate measuring at 42 mAs was evident. Measuring anterior teeth
(especially in the mandible) and dehiscences the real bone level tended to be
underestimated, single measuring points showed great deviations. There was a
good reliability of detection dehiscences (sensitivity 97/94 %, specifity 98
%), in contrast to fenestrations which were mostly overestimated and often
(especially maxilla) detected false-positive (sensitivity 100 %, specifity 86
%). For a resolution of 0,3 mm voxel size areas with a bone thickness less
than 0,39 mm could not be detected properly. The measurements showed
significantly better results for 0,15 mm than 0,3 mm voxel size. Finally,
comparing the results to the radiographic images, a subjectively diffuse
depicted root surface could be a hint of false-positive detection of
dehiscences or fenestrations in this area. Conclusion: Cone-beam computed
tomography provides a good imaging of periodontal structures in three planes,
true to scale and without overlap. For specific clinical consideration it
seems to provide sufficiently precise radiological diagnostics of the osseous
morphology at 0,3 mm voxel size for most discrepancies could be classified as
negligibly small. There are limitations if the bony thickness ranges near to
the spatial resolution of the CBCT leading to false-positive diagnosis of bone
defects. Therefore use of high resolution CBCT is advisable for periodontal
issues.
en
dc.rights.uri
http://www.fu-berlin.de/sites/refubium/rechtliches/Nutzungsbedingungen
dc.subject
cone-beam computed tomography (CBCT)
dc.subject
periodontal defects
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit
dc.title
Einfluss klinischer und bildgebender Parameter auf die Genauigkeit der
Darstellung knöcherner parodontaler Strukturen in der digitalen
Volumentomographie
dc.contributor.firstReferee
N.N.
dc.contributor.furtherReferee
N.N.
dc.date.accepted
2015-02-27
dc.identifier.urn
urn:nbn:de:kobv:188-fudissthesis000000098151-9
dc.title.translated
Influence of clinical aspects and radiological parameters on the imaging of
alveolar bone morphology using cone-beam computed tomography