dc.description.abstract
The most widely used quality control assay for CD34 + hematopoietic stem cell product characterization is the protocol established by the International Society of Hematotherapy and Graft Engineering (ISHAGE). While this protocol is still the gold standard for stem cell enumeration and viability assessment, it does not include T cell enumeration, which is nowadays mandatory for assaying standard allogeneic grafts and various advanced therapy medicinal products (ATMPs). In accordance, we have developed and extensively validated a new approach for a more comprehensive characterization of hematopoietic cellular products using a pre-formulated dried antibody format panel. In addition to the counting beads, the typical markers CD45 fluorescein isothiocyanate (FITC) and CD34 phycoerythrin (PE), as well as the viability dye 7-amino actinomycin D (7-AAD), our novel pre-formulated panel also contains CD3 Pacific Blue (PB) and CD19 allophycocyanin (APC) in the same tube, thereby allowing a combined calculation of leucocytes, stem cells, T and B cells. Showing high linearity, sensitivity and accuracy, our approach is easy to implement and enables a more in-depth characterization of the cellular product under release testing conditions. In addition, the dried pre-formulated antibody approach increases assay reliability compared to the standard antibody panel.
en
dc.rights.uri
https://creativecommons.org/licenses/by/4.0/
dc.subject
Hematopoietic Stem Cells
en
dc.subject
Antigens, CD34
en
dc.subject
Flow Cytometry
en
dc.subject
Phycoerythrin
en
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit::610 Medizin und Gesundheit
dc.title
Widely applicable, extended flow cytometric stem cell enumeration panel for quality control of advanced cellular products
dc.type
Wissenschaftlicher Artikel
dcterms.bibliographicCitation.articlenumber
17995
dcterms.bibliographicCitation.doi
10.1038/s41598-022-22339-1
dcterms.bibliographicCitation.journaltitle
Scientific Reports
dcterms.bibliographicCitation.number
1
dcterms.bibliographicCitation.originalpublishername
Springer Nature
dcterms.bibliographicCitation.volume
12
refubium.affiliation
Charité - Universitätsmedizin Berlin
refubium.funding
Springer Nature DEAL
refubium.resourceType.isindependentpub
no
dcterms.accessRights.openaire
open access
dcterms.bibliographicCitation.pmid
36289245
dcterms.isPartOf.eissn
2045-2322