dc.description.abstract
Hintergrund
Die Synchronisation des kardialen Zyklus mit der Bildakquise (kardiales Gating) spielt in der kardiovaskulären Magnetresonanztomographie (CMR) eine große Rolle. Das dafür häufig verwendete Elektrokardiogram (EKG) unterliegt in der Umgebung des Magnetresonanztomographen häufig multiplen Störfaktoren. Insbesondere die Parameter Magnetfeldstärke, verwendete Gating-Methode und Positionierung der Elektroden scheinen die Güte des Gatings zu beeinflussen.
Deren quantitativer Einfluss ist bis jetzt nicht hinreichend erforscht. Diesen abschätzen zu können, wäre jedoch für die Planung von Studien und die Evaluierung von neu entwickelten Gating-Methoden von großem Interesse. Daher war es Ziel dieser Studie, den Einfluss von 1.5T, 3T und insbesondere auch ultra-hohen Feldstärken von 7T, der EKG- und Vektorkardiogramm(VKG)-basierten Gating-Methode sowie von vier standardisierten Elektrodenpositionen auf die Güte des Gatings zu untersuchen.
Methoden
EKG-Signale von 16 gesunden Proband*innen wurden bei Feldstärken von 1.5T, 3T und 7T, jeweils mit zwei verschiedenen Gatingmethoden und vier standardisierten Elektrodenpositionen in einem standardisierten CMR-Protokoll aufgenommen.
Die quantitative Güte des Gatings wurde mittels spezialisierter Software analysiert, wobei eine manuelle Annotation der QRS-Komplexe als Goldstandard zur Berechnung der Metriken für die Gating-Güte diente. Die statistische Analyse erfolgte in einem dreistufigen Testmodell.
Ergebnisse
In der globalen Analyse konnte ein signifikanter Einfluss der Magnetfeldstärke für alle untersuchten Gütemetriken gezeigt werden, wobei ultra-hohe Feldstärken zu einer deutlichen Verminderung der Gating-Güte führten (p<0.0001). Für die untersuchten Gating-Methoden zeigte sich in unseren Daten insgesamt eine Überlegenheit der EKG-Methode gegenüber der VKG-Methode für die Metriken: Sensitivität: p<0.0001,
positiv prädiktiver Wert: p=0.0005, Propagation Delay: p>0.0001 und Jitter: p>0.0001
Die untersuchten alternativen Elektrodenpositionen zeigten in der Subgruppenanalyse eine ähnliche Gating-Güte wie die Standardpositionen.
Zusammenfassung
Zusammenfassend konnte der Einfluss von Magnetfeldstärken von 1.5T, 3T und 7T, der EKG- und VKG-Gating-Methode und von vier standardisierten Elektrodenposition auf die Gating-Güte quantifiziert werden. Die Ergebnisse erlauben eine verbesserte Abschätzung der Abbruchraten und Gesamtmesszeiten in CMR-Studien sowie eine Optimierung des Gatings für verschiedene Messkonfigurationen.
de
dc.description.abstract
Purpose
Synchronizing the cardiac cycle with image acquisition (cardiac gating) is of paramount importance in cardiovascular magnetic resonance (CMR). Inside the scanner the electrocardiogram (ECG) generally utilized in gating is often compromised by several factors. Magnetic field-strength, gating method and electrode positioning seem to influence gating performance. Their quantitative impact, however, is not yet fully understood. An estimation of this impact, however, could facilitate study design and evaluation of newly developed gating methods. Therefore, the aim of this study was to systematically assess and quantify the impact of field-strengths of 1.5T, 3T and 7T, the ECG-based and the vectorcardiogram(VCG)-based gating-method, as well as four standardized electrode positions on gating performance.
Methods
ECG-Signals of 16 healthy volunteers were recorded at field-strengths of 1.5T, 3T and 7T, with the ECG-method and the VCG-method and with four standardized electrode positions during a standardized CMR-protocol.
Gating-performance was assessed and quantified via specialized software with manual annotation of the QRS-complexes serving as the reference standard for the calculation of performance metrics. The statistical analysis was performed in a hierarchical test-strategy with three levels.
Results
In the global analysis a significant impact of magnetic field strengths on gating performance was found for all performance metrics, with ultra-high field-strengths leading to a significant decrease in gating performance (p<0.0001). In our data the ECG-method provided a better performance than the VCG-method overall. For sensitivity: p<0.0001, positive predictive value: p=0.0005, propagation delay: p>0.0001 and jitter: p>0.0001. In the subgroup-analysis, the assessed alternative electrode positions provided similar gating-performance to the standard-positions.
Conclusion
We successfully quantified the influence of 1.5T, 3T and 7T, the ECG- and the VCG-method and four standardized electrode positions on gating-performance. Our results allow an improved estimation of drop-out rates and scan durations for CMR-studies, as well as an optimization of the gating procedure for specific gating-configurations.
en
dc.rights.uri
https://creativecommons.org/licenses/by/4.0/
dc.subject
Ultra-High-Field
en
dc.subject
Cardiac Gating
en
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit::610 Medizin und Gesundheit
dc.title
Evaluierung und Einfluss von Gating-Methoden für klinische Feldstärken und Ultra-Hochfeld-Magnetresonanztomographie
dc.contributor.gender
male
dc.contributor.firstReferee
N.N.
dc.contributor.furtherReferee
N.N.
dc.date.accepted
2024-03-23
dc.identifier.urn
urn:nbn:de:kobv:188-refubium-41457-0
dc.title.translated
Evaluation and impact of gating methods for clinical and ultra-high-field magnetic resonance imaging
eng
refubium.affiliation
Charité - Universitätsmedizin Berlin
dcterms.accessRights.dnb
free
dcterms.accessRights.openaire
open access
dcterms.accessRights.proquest
accept