Objectives: Detectability experiments performed to assess the diagnostic performance of computed tomography (CT) images should represent the clinical situation realistically. The purpose was to develop anatomically realistic phantoms with low-contrast lesions for detectability experiments.
Methods: Low-contrast lesions were digitally inserted into a neck CT image of a patient. The original and the manipulated CT images were used to create five phantoms: four phantoms with lesions of 10, 20, 30, and 40 HU contrast and one phantom without any lesion. Radiopaque 3D printing with potassium-iodide-doped ink (600 mg/mL) was used. The phantoms were scanned with different CT settings. Lesion contrast was analyzed using HU measurement. A 2-alternative forced choice experiment was performed with seven radiologists to study the impact of lesion contrast on detection accuracy and reader confidence (1 = lowest, 5 = highest).
Results: The phantoms reproduced patient size, shape, and anatomy. Mean ± SD contrast values of the low-contrast lesions were 9.7 ± 1.2, 18.2 ± 2, 30.2 ± 2.7, and 37.7 ± 3.1 HU for the 10, 20, 30, and 40 HU contrast lesions, respectively. Mean ± SD detection accuracy and confidence values were not significantly different for 10 and 20 HU lesion contrast (82.1 ± 6.3% vs. 83.9 ± 9.4%, p = 0.863 and 1.7 ± 0.4 vs. 1.8 ± 0.5, p = 0.159). They increased to 95 ± 5.7% and 2.6 ± 0.7 for 30 HU lesion contrast and 99.5 ± 0.9% and 3.8 ± 0.7 for 40 HU lesion contrast (p < 0.005).
Conclusions: A CT image was manipulated to produce anatomically realistic phantoms for low-contrast detectability experiments. The phantoms and our initial experiments provide a groundwork for the assessment of CT image quality in a clinical context.