dc.contributor.author
Reshetnik, Alexander
dc.contributor.author
Compton, Friederike
dc.contributor.author
Schoelzel, Anna
dc.contributor.author
Toelle, Markus
dc.contributor.author
Zidek, Walter
dc.contributor.author
van der Giet, Markus
dc.date.accessioned
2018-06-08T10:32:45Z
dc.date.available
2017-10-10T09:35:13.696Z
dc.identifier.uri
https://refubium.fu-berlin.de/handle/fub188/20624
dc.identifier.uri
http://dx.doi.org/10.17169/refubium-23925
dc.description.abstract
Assessment of the cardiac output (CO) is usually performed with invasive
techniques requiring specialized equipment in the intensive care unit (ICU).
With TEL-O-GRAPH (TG), CO can be derived from the oscillometrically obtained
brachial pulse wave during the measurement of brachial blood pressure. CO and
stroke volume (SV) determinations with TG were compared with transpulmonary
thermodilution measurements with the PICCO system (PICCO) in 38
haemodynamically unstable ICU patients with a total of 84 comparison
measurements performed. SV (33.3 ± 9.0 ml/m2 vs. 44.3 ± 14.4 ml/m2, p < 0.001)
and CO (2.7 ± 0.5 l/min/m2 vs. 3.8 ± 1.2 l/min/m2, p < 0.001) were
underestimated significantly with TG and oscillometric brachial systolic blood
pressure (BP) was significantly lower and diastolic BP significantly higher
than invasive femoral artery pressure. A linear correlation was found between
CO dimension and CO underestimation with TG. Correct tracking of CO changes
with a fluid challenge was possible in 69.5% of measurements. Oscillometric
noninvasive CO is possible in the ICU, but accuracy and precision of this new
method are lacking. Implementation of a correction factor accounting for the
linear increase in CO underestimation observed with increasing CO could
improve CO assessment with TG in haemodynamically unstable patients.
en
dc.rights.uri
http://creativecommons.org/licenses/by/4.0/
dc.subject
Cardiovascular diseases
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit
dc.title
Noninvasive oscillometric cardiac output determination in the intensive care
unit – comparison with invasive transpulmonary thermodilution
dc.type
Wissenschaftlicher Artikel
dcterms.bibliographicCitation
Scientific Reports. - 7 (2017), Artikel Nr. 9997
dcterms.bibliographicCitation.doi
10.1038/s41598-017-10527-3
dcterms.bibliographicCitation.url
http://www.nature.com/articles/s41598-017-10527-3
refubium.affiliation
Charité - Universitätsmedizin Berlin
de
refubium.mycore.fudocsId
FUDOCS_document_000000028233
refubium.note.author
Der Artikel wurde in einer reinen Open-Access-Zeitschrift publiziert.
refubium.resourceType.isindependentpub
no
refubium.mycore.derivateId
FUDOCS_derivate_000000008895
dcterms.accessRights.openaire
open access