dc.contributor.author
Balzer, Felix
dc.contributor.author
Habicher, Marit
dc.contributor.author
Sander, Michael
dc.contributor.author
Sterr, Julian
dc.contributor.author
Scholz, Stephanie
dc.contributor.author
Feldheiser, Aarne
dc.contributor.author
Mueller, Michael
dc.contributor.author
Perka, Carsten
dc.contributor.author
Treskatsch, Sascha
dc.date.accessioned
2018-06-08T03:55:40Z
dc.date.available
2016-09-20T07:59:54.171Z
dc.identifier.uri
https://refubium.fu-berlin.de/handle/fub188/16232
dc.identifier.uri
http://dx.doi.org/10.17169/refubium-20416
dc.description.abstract
Objective Continuous invasive arterial blood pressure (IBP) monitoring remains
the gold standard for BP measurement, but traditional oscillometric non-
invasive intermittent pressure (NIBP) measurement is used in most low-to-
moderate risk procedures. This study compared non-invasive continuous arterial
BP measurement using a Nexfin® monitor with NIBP and IBP monitors. Methods
This was a single-centre, prospective, pilot study in patients scheduled for
elective orthopaedic surgery. Systolic BP, diastolic BP and mean arterial
blood pressure (MAP) were measured by Nexfin®, IBP and NIBP at five
intraoperative time-points. Pearson correlation coefficients, Bland–Altman
plots and trending ability of Nexfin® measurements were used as criteria for
success in the investigation of measurement reliability. Results A total of 20
patients were enrolled in the study. For MAP, there was a sufficient
correlation between IBP/Nexfin® (Pearson = 0.75), which was better than the
correlation between IBP/NIBP (Pearson = 0.70). Bland–Altman analysis of the
data showed that compared with IBP, there was a higher percentage error for
MAPNIBP (30%) compared with MAPNexfin® (27%). Nexfin® and NIBP underestimated
systolic BP; NIBP also underestimated diastolic BP and MAP. Trending ability
for MAPNexfin® and MAPNIBP were comparable to IBP. Conclusion Non-invasive BP
measurement with Nexfin® was comparable with IBP and tended to be more precise
than NIBP.
en
dc.rights.uri
http://creativecommons.org/licenses/by-nc/3.0/
dc.subject
Blood pressure
dc.subject
non-invasive arterial pressure
dc.subject
invasive arterial pressure
dc.subject
Nexfin® Trial registration: ClinGov. NCT01263990
dc.subject.ddc
600 Technik, Medizin, angewandte Wissenschaften::610 Medizin und Gesundheit
dc.title
Comparison of the non-invasive Nexfin® monitor with conventional methods for
the measurement of arterial blood pressure in moderate risk orthopaedic
surgery patients
dc.type
Wissenschaftlicher Artikel
dcterms.bibliographicCitation
Journal of International Medical Research. - 44 (2016), 4, S. 832-843
dcterms.bibliographicCitation.doi
10.1177/0300060516635383
dcterms.bibliographicCitation.url
http://imr.sagepub.com/content/44/4/832
refubium.affiliation
Charité - Universitätsmedizin Berlin
de
refubium.mycore.fudocsId
FUDOCS_document_000000025396
refubium.note.author
Der Artikel wurde in einer Open-Access-Zeitschrift publiziert.
refubium.resourceType.isindependentpub
no
refubium.mycore.derivateId
FUDOCS_derivate_000000006984
dcterms.accessRights.openaire
open access