LC–MS is one of the most important tools for the comprehensive characterization of N-glycans. Despite many efforts to speed up glycan analysis via optimized sample preparation (e.g., faster enzyme digestion in combination with instant or rapid labeling dyes), a major bottleneck remains the rather long measurement times of HILIC chromatography. Further complication arises from the necessity to concomitantly calibrate with an external standard to allow for accurate retention times and the conversion into more robust GU values. Here we demonstrate the use of an internal calibration strategy for HILIC chromatography to speed up glycan analysis. By reducing the number of utilized dextran oligosaccharides, the calibrant can be spiked directly into the sample such that external calibration runs are no longer required. The minimized dextran ladder shows accurate GU calibration with a minor deviation of well below 1% and can be applied without modifications in sample preparation or data processing. We further demonstrate the simultaneous use of the minimized dextran ladder as calibrant for the estimation of CCS values in traveling wave ion mobility spectrometry. In both cases, the minimized dextran ladder enables the measurement of calibrant and sample in a single HPLC run without losing information or accuracy.