Title:
Temperature stability of the sky quality meter
Author(s):
Schnitt, Sabrina; Ruhtz, Thomas; Fischer, Jürgen; Hölker, Franz; Kyba, Christopher C. M.
Year of publication:
2013
Available Date:
2013-12-28T11:54:19.785Z
Abstract:
The stability of radiance measurements taken by the Sky Quality Meter (SQM)was
tested under rapidly changing temperature conditions during exposure to a
stable light field in the laboratory. The reported radiance was found to be
negatively correlated with temperature, but remained within 7% of the initial
reported radiance over a temperature range of -15 °C to 35 °C, and during
temperature changes of -33 °C/h and +70 °C/h.This is smaller than the
manufacturer’s quoted unit-to-unit systematic uncertainty of 10%,indicating
that the temperature compensation of the SQM is adequate under expected
outdoor operating conditions.
Keywords:
light pollution
sky quality meter
light at night
temperature
stability
skyglow
night sky brightness
DDC-Classification:
520 Astronomie
Publication Type:
Wissenschaftlicher Artikel
Also published in:
Sensors 2013, 13 (9)
URL of the Original Publication:
DOI of the Original Publication:
Department/institution:
Geowissenschaften
Institut für Meteorologie / Institut für Weltraumwissenschaften
Comments:
Gefördert durch die DFG und den Open Access Publikationsfonds der Freien
Universität Berlin